|
Laser-Induced Fluorescence
Description
Laser-induced Fluorescence (LIF) is a method for real-time, in-situ, field screening of hydrocarbons
in undisturbed subsurface soils and groundwater. The technology is intended
to provide highly detailed, qualitative to semiquantitative information about
the distribution of subsurface petroleum contamination. LIF sensors are deployed
as part of integrated mobile cone penetrometer (CPT) systems that are operated
by highly trained technicians familiar with the technology and its application.
Typical Uses
LIF can detect gasoline, diesel fuel, jet fuels, fuel oil, motor oil, grease, and coal tar
in the subsurface. This screening data can be used to guide an investigation
or removal action or to delineate the boundaries of a subsurface contamination
plume prior to installing monitoring wells or taking soil samples.
There are currently two major LIF systems available, the SCAPS and Rapid Optical Screening Tool
(ROST) systems. The Site Characterization and Analysis Penetrometer System
(SCAPS) LIF system is one of several CPT-mounted sensors developed through a
collaborative effort of the Army, Navy, and Air Force under the Tri-Services
Program. The ROST system was developed by Loral Corporation and Dakota
Technologies, Inc. The SCAPS LIF is available only through the U.S. Army Corps
of Engineers (USACE) and the US Navy. ROST is available commercially through
Fugro, Inc. Both systems, while differing in some respects, are very similar
in their theories and methods of operation.
Click here to see a brief comparison of the two systems.
Theory
of Operation
The method uses
a fiber optic-based LIF sensor system deployed
with a standard 20-ton cone penetrometer (CPT). Light at a specific wavelength
generated from a laser is passed down a fiber optic cable to a sapphire window
in the tip of the CPT rod string as it is advanced
into the subsurface. The laser light excites two- or three-ring aromatic compounds,
or polycyclic aromatic hydrocarbons (PAH), in the soil adjacent to the sapphire
window, causing them to fluoresce. The relative response of the sensor depends
on the specific analyte being measured because of the varying ratios of PAHs
in each hydrocarbon mixture. The induced fluorescence from the PAHs is returned
over a second fiber to the surface where it is quantified using a detector system.
The peak wavelength and intensity provide information about petroleum product
type or potential interferences. The intensity of the fluorescence is used as
an indicator of the relative contaminant concentration.
System Components
Both the SCAPS
and the ROST sensing devices are part of an integrated system consisting
of a cone penetrometer (CPT) truck designed for direct push sampling, the laser,
and a detector. The LIF sensor is used in conjunction with the stratigraphic
sensing devices contained on the CPT truck. CPT and standard penetrometer testing
are widely used in the geotechnical industry for determining soil strength and
soil type from measurements of tip resistance and sleeve friction on an instrumented
probe.
The
forward portion of the truck-mounted laboratory is the push room. It contains
the rods, hydraulic rams, and associated system controllers. The weight of the
CPT truck provides a 20-ton static reaction force for advancing the probe rod
into the ground, which allows the hydraulics to advance threaded-end steel rod
into the ground at about 1 meter per minute. The rods, sensing probes, and sampling
tools can be advanced to 50 meters or more in soil. As the rods are withdrawn,
grout can be injected through probe rods to seal the push hole. The trucks are
also fitted with a self-contained decontamination system that allows the rods
and probe to be steam cleaned as they are withdrawn from the push hole, through
the steam cleaning manifold, and back into the CPT push room.
The rear portion
of the truck contains the laboratory in which components of the LIF
system and onboard computers are located. The primary differences between
the two systems are the laser and the detector systems. The SCAPS
LIF system uses a pulsed nitrogen laser that produces light at a fixed
wavelength of 337 nanometers (nm), which is passed down a fiber optic cable
inside the probe rods to the sapphire window in the sensor tip of the CPT rod
string. Fluorescence stimulated in the in situ soil “sample” by the
laser is collected by a second fiber and returned to an optical multichannel
analyzer (OMA) spectrograph, where it is dispersed spectrally on a photodiode
array (PDA) detector. This arrangement allows for the rapid acquisition of spectral
data, so that as the SCAPS LIF sensor is pushed into the soil, real-time plots
are generated of depth versus maximum fluorescence intensity and the wavelength
at which the maximum intensity occurs.
By contrast, the
ROST system uses a tunable dye laser that can produce light at variable
wavelengths chosen by the operator, typically 290 NM The detection system for
the ROST consists of a monochromator, a photomultiplier tube, and digital
storage oscilloscope (DSO). The monochromator acts as a variable wavelength
narrow bandpass filter. By acquiring fluorescence data at a series of wavelengths,
the fluorescence technician can determine the wavelength of maximum intensity
in the fluorescence spectrum. The light passing through the monochromator at
this wavelength is then converted to an electrical signal by the photomultiplier.
The signal from the photomultiplier is fed to the DSO, which displays the waveform
plot (fluorescence intensity as a function of time following the excitation
laser pulse).
Mode
of Operation
Because of both
the complexity of the LIF methods and the specialized requirements of operating
a CPT, the operation of either the SCAPS LIF system or the ROST takes
considerable experience. For this reason, both are designed to be operated by
their own trained technicians, not the general public. The SCAPS CPT and LIF
technology is not available for use by private citizens or corporations, but
is available to state and federal agencies. The ROST is available commercially
through the Fugro corporation.
Both systems are
typically deployed with a three-person crew and a geologist. Two people are
needed to handle the push rods and operate the hydraulic press, and the third
person operates the sensor, including measurements of the calibration and control
standards, and monitoring the real-time CPT geotechnical data and fluorescence
response from the soils. Once the system has been calibrated by the operator,
the CPT truck is set up over the designated location for a push. Continuing
calibration checks may be performed using a calibration standard held against
the sapphire window before and after each push.
For
the SCAPS LIF system, a qualitative identification of different types of petroleum
products can be gathered from plots of fluorescence intensity versus wavelength.
Under normal operating conditions, fluorescence emission spectra are collected
once per second as the penetrometer probe is pushed into the ground at a rate
of approximately 1 meter per minute. This yields a measurement with a vertical
spatial resolution of approximately 0.2 feet. A computer equipped with custom
software controls the fiber optic fluorometer sensor system and stores fluorescence
emission spectra and conventional CPT sleeve friction and tip resistance data.
The computer also generates real-time depth plots of fluorescent intensity at
the spectral peak, wavelength of spectral peak, sleeve friction and tip resistance,
and soil type characteristics as interpreted from the CPT data. The fluorescent
intensity in the spectral window is plotted
as a function of depth in real time as the probe is pushed into the
soil, creating a semiquantitative representation of the subsurface contamination.
The entire fluorescent emission spectrum is also stored on a fixed hard disk
for post-processing or comparison with confirmatory data.
The
ROST can be operated in two modes. In the dynamic mode, the fluorescence
intensity is plotted as a function of depth below ground surface in
real-time while the sensor is being advanced in the subsurface. The ROST
system operator chooses the excitation laser wavelength and fluorescence emission
monitoring wavelength and they are held constant during the push. Once areas
of significant contamination have been tentatively identified in the dynamic
mode, the CPT is held at a fixed depth and the ROST can be operated in
the static mode to identify the general class of contamination present. During
the static mode, ROST can obtain wavelength-time-matrices
(WTM) which represent a 3-dimensional plot of relative fluorescence
intensity versus fluorescence lifetime versus wavelength. WTMs produce contaminant-class-specific
three-dimensional figures that can be used to identify the type of fuel that
is present by comparing them with an on-board library of WTM “fingerprints”
from common contaminants. Normally, in the static mode, the excitation wavelength
is held constant and the emission monitoring wavelength is varied.
As the rods are
withdrawn, grout is injected through probe rods to seal the push hole. The rods
and probe are steam cleaned as they are withdrawn from the push hole and back
into the CPT push room.
An SOP is available
by clicking here.
Target
Analytes
LIF can detect
gasoline, diesel fuel, jet fuels, fuel oil, motor oil, grease, and coal tar
in the subsurface.
Performance
Specs
Interferences
The in situ
fluorescence response of the LIF sensor to hydrocarbon compounds is sensitive
to a number of interferences, but variations in the soil matrix are the most
pronounced. LIF sensitivity to petroleum hydrocarbons in soil has been shown
to be inversely proportional to the available surface area of the soils. Sandy
soils tend to have a much lower total available surface area than clay soils,
so hydrocarbon compounds in sandy soils generally yield a higher fluorescence
response than they do in clay-rich soils.
Although intended
to specifically target petroleum hydrocarbons, the excitation energy produced
by the LIF system's laser may cause other substances to fluoresce as well, which
may cause interference problems. Many common fluorescent minerals such as calcite
can produce a measurable LIF signal. Other man-made non-hydrocarbon fluorescent
material may be found in the subsurface environment: deicing agents, antifreeze
additives, and many detergent products are all known to fluoresce very strongly,
for example. Naturally occurring organic matter, which include PAHs, also can
fluoresce. In many cases it is possible for an experienced operator to differentiate
between the fluorescent signatures of hydrocarbons and other interfering compounds.
Detection Limits
LIF data quality
is sufficient for qualitative screening, and relative intensities may be considered
quantitative screening level data only. Site-specific detection limits vary
from levels of 50 to 1,000 mg/kg, but exact detection limits are difficult to
determine and will vary between sites and petroleum products. For example, according
to results published in an EPA Innovative Technology Verification Report, the
SCAPS LIF detection threshold is approximately 100 to 300 mg/kg for total petroleum
hydrocarbons as confirmed by EPA Method 418.1
The effective upper
detection range of both LIF detectors depend on the specific hydrocarbon analyte
as well as the particular matrix. Generally, the response curves generated during
calibration remain linear until approximately 10,000 mg/kg, when the response
trails off. The upper effective range may be extended to higher concentrations
by the operator, but this results in decreased sensitivity at lower concentrations.
Calibration
LIF systems measure
the relative intensity of fluorescence in soils caused by hydrocarbon contamination.
It is critical that these measurements be accurate if the data is to be useful
for project decision making. For this reason, both systems must be calibrated
prior to use.
The SCAPS LIF sensor
is calibrated using spiked soil samples representative of the site. Diesel fuel
marine standard or other petroleum hydrocarbons with a fluorescence response
appropriate for the site are used to spike the soil samples. The ROST
system is calibrated with a proprietary blend of synthetic motor oil and other
substances. In both cases, the calibration standards are run in triplicate at
the beginning of each day and again if equipment is changed. After measurement,
the average and standard deviation is computed for each sample and the sample
is rerun if the standard deviation exceeded 20 percent. A calibration
curve is generated by plotting the average of maximum fluorescence peak
intensity versus the concentration of fuel product added to the calibration
soil sample.
Sample Preparation
Because LIF is
deployed on a direct push CPT rod and analysis is conducted directly in situ,
no sample preparation is possible. Some site preparation may be required, however.
Hard surfaces such as concrete or caliche may require drilling or cutting prior
to advancing the CPT rod into the ground.
Quality Control
Even though they
are not quantitative systems, the data generated by the LIF system must be of
a known and acceptable quality if it is to be useful for project decision making.
For this reason, it is critical that the quality of the data produced by the
LIF system be determined and documented. There are several types of quality
control checks that can be applied to assess whether the LIF systems are functioning
properly and are producing accurate and useable data.
The SCAPS LIF sensor's
response is checked using a quinine sulfate fluorescent standard before and
after each push. This measurement is a check of system performance and provides
a means for normalizing measurements. If the fluorescent intensity changed by
more than 20 percent of the initial value determined during pre-push calibration,
system trouble shooting procedures are initiated.
A system check
using a reference solution is performed before and after each ROST push
as well. The reference is a selected mixture of hydrocarbons in solution contained
in a standard fluorescence cuvette that can be strapped onto the sensor tip
outside the sapphire window. Both wavelength and intensity of the standard are
monitored. If the wavelength differs by greater than 5 NM from the known value,
a wavelength calibration is performed. If the intensity changes by more than
20 percent, system troubleshooting is required.
A clean sand blank
may be measured pre- and post-push as part of the standard data collection procedure.
The blank helps assure that the sapphire window does not become contaminated
and that the sensor doesn't develop a “memory effect” from previous samples.
If the clean sand blank LIF measurement varies beyond 50 percent of its pre-push
calibration value, troubleshooting procedures must be initiated.
Finally, a qualitative
assessment can be made by comparing subsurface contaminant cross sections generated
from either SCAPS LIF or ROST data to borehole logs or cross sections
prepared using conventional methods such as a hollow stem auger rig and sampling
data generated using EPA-approved analytical methods.
Precision and Accuracy
Precision refers to the reproducibility of measurements of the same characteristic, usually
under a given set of conditions. Accuracy refers to the degree of agreement of
a measurement to the “true” value, as determined by traditional analytical methods.
Both provide a measure of the LIF system's performance and can help determine
how useful their data is.
Precision is usually
assessed by comparing the results of duplicate analyses. However, because both
LIF sensors' are in situ sensors, it is not possible to obtain true duplicate
analyses. Instead, an estimate of the instrumental precision can be obtained
for either system by evaluating the results from multiple measurements of their
respective calibration check samples, which are analyzed before and after each
push. During an ETV verification study of the SCAPS, the standard deviation
of 20 check sample measurements was less than 1 percent of the mean count. The
standard deviation of 20 check sample measurements during a corresponding ROST
demonstration was 2.2% of the mean count.
Because neither
LIF system provides quantitative results, accuracy is assessed qualitatively
by measuring the agreement between “detect/nondetect” determinations made by
the LIF and corresponding confirmatory laboratory samples. For example, if the
laboratory result was above the LIF's detection limit and the average LIF data
from the push at the corresponding depth exceeded the LIF fluorescence threshold,
the results agree. If the average LIF data were below the threshold and the
corresponding analytical data were above the corresponding detection threshold,
the result was a “false negative,” which is the most serious error in terms
of environmental sampling. The false negative rate for the SCAPS during the
ETV demonstration was 4.9 percent. At least 90 percent of the samples analyzed
during the ETV demonstration of the ROST agreed with the confirmatory
results, and the false negative rate ranged from 3.3 percent to 10 percent,
depending on the confirmatory method used.
Superfund
Innovative Technology Evaluation (SITE) Demonstration
The SCAPS LIF and
ROST systems were evaluated in 1994 during a demonstration conducted under
EPA's SITE program. The two systems were demonstrated at three sites in the
Midwest that had varying concentrations of coal tar waste and petroleum fuels,
and a range of soil textures. A qualitative assessment was made by comparing
subsurface contaminant cross sections generated from either SCAPS LIF or ROST
data to cross sections prepared using conventional methods (hollow stem auger
rig and EPA-approved analytical methods) for the three sites. For both systems,
the chemical cross sections were comparable to the conventional methods. The
chemical cross sections for both systems showed close agreement to the conventional
method cross sections in identifying low, medium, and high zones of contamination.
Generally, the relative fluorescence intensity was positively related to the
total petroleum hydrocarbon (TPH) and total PAH concentrations. However, a good,
quantitative correlation between the fluorescence intensity and individual or
classes of petroleum compounds could not be found for either system. In only
one case during the demonstration did either system produce a false negative
where conventional samples indicated contamination in the hundreds of ppm range.
The results of the SITE demonstration for both systems can be found in individual
Innovative Technology Evaluation Reports.
State of
California Certification
Technology field
validation studies at nine sites were conducted for the state of California
for the SCAPS LIF system. Between 16 and 45 CPT pushes, along with 3 to 8 confirmation
soil sample borings, were completed at each site. For the 164 TPH analyses completed,
there were 9, or 5.5 percent, false positives and 12, or 7.3 percent, false
negatives. For the 164 total recoverable petroleum hydrocarbon analyses, there
were 6, or 3.7 percent, false positives and 16, or 9.8 percent, false negatives.
The California
Military Environmental Coordination Committee (CMECC) guidance lists LIF as
a screening tool and indicates that it should not be used to generate definitive
data.
Advantages
The primary advantage
of using LIF systems is their ability to provide real-time chemical and geological
information while in the field. This data can reduce and focus the amount of
physical sampling and laboratory analysis, as well as optimize monitoring well
placement.
Both systems are
capable of achieving 200 to 300 feet of pushes in a 10-hour work day.
The vertical spatial
resolution is near 2.0 cm, which allows small zones of contamination to be delineated
that might be missed by conventional sampling protocols.
No drill cuttings
are produced with the system, saving the logistical requirement of handling
drums of cuttings and eliminating disposal costs.
The sample holes
can be grouted as the push rod is pulled from the hole. Also, the push rod can
be decontaminated remotely as it is retracted from the hole. All the decontamination
fluids are containerized in the process.
During a SITE demonstration,
three sites were characterized using the SCAPS LIF system and the ROST
system at a cost of $20,000 and $41,200, respectively. These costs can be compared
to the approximate $43,000 using conventional drilling methods and on-site analytical
capabilities. Both systems cost less than the reference methods and produced
almost 1,200 more data points in a real-time fashion, as opposed to waiting
hours or days for the data.
Limitations
The operation of
the SCAPS LIF system or the ROST system takes considerable experience.
It takes many days and numerous projects to become familiar with the operation
of the technology. They are provided as services by their respective vendors
for this reason.
Although these
sensors provide a relative degree of contamination that closely matches reference
method data, little direct, quantitative correlation has been found to individual
or classes of petroleum compounds.
The cost of one
of these systems may be prohibitive on small-scale projects. They have primarily
been used at large sites such as Department of Defense (DoD) and Department
of Energy (DOE) facilities.
Some maintenance
of the CPT tools and the LIF sensors is required and breakdowns can be expected
on long-term projects. Downtime due to breakage of fiber optic cables and push
rods, fogging of the sapphire window, and problems with the grout pump or decontamination
unit may occur.
These systems can
only be used where direct push is feasible, such as in unconsolidated sediments,
and where a 20-ton CPT truck can gain access.
The sensors are
limited to a depth of 50 meters because of attenuation in the optical fiber
umbilical cord.
Minerals such as
calcite, naturally occurring organic matter, and man-made chemicals also can
fluoresce, which may cause interference problems.
Work performed
by the ROST system at Dover Air Force Base indicated the potential for
smearing and a memory effect on the sensor.
Cost Data
The estimated cost
of sampling using the SCAPS LIF system varies between $12 and $20
per foot depending upon whether the operators provide a turnkey operation or
the customer provides field deployment assistance such as permitting, site management,
and development of work and health and safety plans. Under normal conditions,
200 to 300 feet of pushes can be advanced per day.
The ROST
technology is available for use within the 48 contiguous United States for a
cost of approximately
$5,300 per day or site-specific footage rates, which includes a CPT rig provided
by a commercial
vendor such as Fugro Geosciences. Crew per diem and mobilization costs are additional
and site specific. Concrete coring, grouting, permit fees, and distant travel
costs or mobilization/demobilization costs vary with each deployment and thus
are not included. This translates to a cost of under $20 per foot.
The main savings
attributable to an LIF system is that it can substantially reduce the number
of monitoring wells drilled at a site. In a large general site characterization
effort, it can provide data in less time and far less expensively than conventional
drilling and sampling. Investigation-derived wastes are minimized along with
the expense of handling and disposal.
During a SITE demonstration,
three sites were characterized using the SCAPS LIF system and the ROST
system at a cost of $20,000 and $41,200, respectively. These costs can be compared
to the approximate $43,000 using conventional drilling methods and on-site analytical
capabilities. Both systems cost less than the reference methods and produced
almost 1,200 more data points in a real-time fashion, as opposed to waiting
hours or days for the data.
The cost of one
of these systems may be prohibitive on small-scale projects. They have primarily
been used at large sites such as Department of Defense (DOD) and Department
of Energy (DOE) facilities.
Manufacturers and
service providers listed below should be contacted directly for cost information.
Verification/Evaluation
Reports
Verification of
the performance of site characterization and field analytical technologies is
conducted through a variety of programs. Evaluation and verification reports
from EPA's Superfund Innovative Technologies Evaluation (SITE) Measuring and
Monitoring Program, EPA's Environmental Technology Verification Program (ETV)
program, along with links to certification statements from California EPA's
(CalEPA) California Environmental Technology Certification Program, are provided
below.
Superfund
Innovative Technologies Evaluation (SITE) Measuring and Monitoring Program
The SITE Demonstration Program encourages the development and implementation
of innovative treatment technologies for (1) remediation of hazardous waste
sites and (2) monitoring and measurement. In the SITE Demonstration Program,
the technology is field-tested on hazardous waste materials. Engineering and
cost data on the innovative technologies are gathered so that potential users
can assess the technology's applicability to a particular site. Data collected
during the field demonstration are used to assess the performance of the technology,
the potential need for pre- and post-treatment processing of the waste, applicable
types of wastes and waste matrices, potential operating problems, and approximate
capital and operating costs. The following reports from the measuring and monitoring
program are available for laser-induced fluorescence:
- This report
describes the demonstration and evaluation of the Site
Characterization and Analysis Penetrometer System (SCAPS) Laser Induced
Fluorescence (LIF) sensor developed by the Tri-Services (Army, Navy, and Air
Force). The SCAPS LIF is designed to provide rapid real-time, relatively low
cost analysis of the physical and chemical characteristics of subsurface soil
to quickly distinguish contaminated areas from noncontaminated areas.
EPA's
Environmental Technology Verification (ETV) Program
EPA's Environmental
Technology Verification (ETV) Program verifies the performance of innovative
technologies. ETV was created to substantially accelerate the entrance of new
environmental technologies into the domestic and international marketplaces.
ETV verifies commercialized, private sector technologies. After the technology
has been tested, the companies will receive a verification report that they
can use in marketing their products. The results of the testing also are available
on the Internet. The following reports from the ETV program are available for
laser-induced fluorescence:
- The Site Characterization
and Analysis Penetrometer System (SCAPS) Laser-Induced Fluorescence (LIF)
Sensor and Support System was verified for in-situ detection of petroleum
hydrocarbons. The verification documents available consist of a verification
report and verification statement.
- Fugro Geosciences, Inc.- Rapid Optical Screening Tool was verified for in-situ detection of petroleum hydrocarbons. The verification documents available consist of a verification report and verification statement.
California
EPA's California Environmental Technology Certification Program
CalEPA's environmental technology certification program
is a voluntary program that provides participating technology developers, manufacturers,
and vendors an independent, recognized third-party evaluation of the performance
of new and mature environmental technologies. Developers and manufacturers define
quantitative performance claims for their technologies and provide supporting
documentation; CalEPA reviews that information and, when necessary, conducts
additional testing to verify the claims. The technologies, equipment, and products
that are proven to work as claimed are given official state certification. The
certification program is voluntary and self-supporting. Companies participating
in the program pay the costs of the evaluation and certification of their technologies.
Technologies that have been certified through this program are listed below.
Links are provided to the web sites that provide the Certified Environmental
Technology Transfer Advisory and Certification Notice for the technologies.
Disclaimer
http://www.clu-in.org/char/technologies/LIF.cfm
Page Last Modified: November 22, 2006

|