|
X-ray Fluorescence
Description
Field-portable, handheld device for simultaneously measuring a
number of metals in various media.
Typical Uses
Energy dispersive X-ray fluorescence (EDXRF) is
a method of detecting metals and other elements, such as arsenic and selenium,
in soil and sediment. Some of the primary elements of environmental concern
that EDXRF can identify are arsenic, barium, cadmium, chromium, copper, lead,
mercury, selenium, silver, and zinc. Field-portable X-ray fluorescence (FPXRF)
units that run on battery power and use a radioactive source were developed
for use in analysis for lead-based paint and now are accepted as a stand-alone
technique for lead analysis. In response to the growing need for field analysis
of metals at hazardous waste sites, many of these FPXRF units have been adapted
for use in the environmental field. The field-rugged units use analytical techniques
that have been developed for analysis of numerous environmental contaminants
in soils. They provide data in the field that can be used to identify and characterize
contaminated sites and guide remedial work, among other applications.
More recently, FPXRF analyzers have been
used to detect metals in water. The water samples must be filtered and concentrated
with an ion exchange membrane to achieve detection limits in the low parts per
billion (ppb) range lower than applicable maximum contaminant levels (MCL).
Many manufacturers of FPXRF units currently are conducting research to refine
the procedures for preparation of water samples to make FPXRF analysis a practical
field analytical technique for metals in water.
Theory of Operation
FPXRF instruments were developed as
an effective and nondestructive tool for measuring lead in paint and in house
dust. Most homes constructed or painted before 1976 contain lead-based paint,
which is one of the most common sources of lead ingested by children. In response,
the U.S. Department of Housing and Urban Development (HUD) set guidelines for
the inspection and abatement of contamination in public housing developments
at which lead paint had been used. HUD considers any paint surface with a lead
content greater than 1 milligram per square centimeter (mg/cm2) to
be a lead-based paint surface. FPXRF units were designed to detect lead in paint
at levels at or lower than that level. Air filters are used to measure concentrations
of metals in household dust. When the volume of air that has passed through
the air filter is measured, a conversion can be made to determine the concentrations
of metals suspended as particulates in the air. Although the technique was developed
for homes in which contamination with lead is suspected, it also has been applied
in monitoring air emissions from industrial processes or from remediation processes
conducted at a hazardous waste site.
One
of the advantages of EDXRF analysis is that it can be used not only to detect
lead, but also to detect and measure many elements simultaneously. Generally,
EDXRF units can detect and quantify elements from atomic number 16 (sulfur)
through 92 (uranium). There are two types of EDXRF units: bench-top units that
use an X-ray tube source and FPXRF analyzers that use a radioisotope as a source
of X-rays. Instruments that use X-ray tubes as sources commonly are not used
in the field because of the larger power requirements for the X-ray tube and
the added weight of the instrument. Use of a radioactive source eliminates the
need for a fixed power source for an X-ray tube, making the FPXRF unit truly
portable.
In
FPXRF analysis, a process known as the photoelectric
effect is used in analyzing samples.
Fluorescent X-rays are produced by exposing a sample to an X-ray source that
has an excitation energy similar to, but greater than, the binding energy of
the inner-shell electrons of the elements in the sample. Some of the source
X-rays will be scattered, but a portion will be absorbed by the elements in
the sample. Because of their higher energy level, they will cause ejection of
the inner-shell electrons. The electron vacancies that result will be filled
by electrons cascading in from outer electron shells. However, since electrons
in outer shells have higher energy states than the inner-shell electrons they
are replacing, the outer shell electrons must give off energy as they cascade
down. The energy is given off in the form of X-rays, and the phenomenon is referred
to as X-ray fluorescence (click to view a schematic
diagram of the X-ray fluorescence process).
Because every element has a different electron shell configuration, each element
emits a unique X-ray at a set energy level or wavelength that is characteristic
of that element. The elements present in a sample can be identified by observing
the energy level of the characteristic X-rays, while the intensity of the X-rays
is proportional to the concentration and can be used to perform quantitative
analysis. In other words, qualitative analysis is performed by observing the
energy of the characteristic X-rays. A quantitative analysis is performed by
measuring the intensity of the X-ray.
The emissions of characteristic X-rays from three
electron shells commonly are involved in FPXRF analysis: the K, L, and M shells.
A typical emission pattern, or emission spectrum, for a given element has several
peaks generated from the emission of X-rays from those shells.
System Components
A FPXRF system has two basic components: the radioisotope
source and the detector. The source irradiates the sample to produce characteristic
X-rays, as described above. The detector measures both the energy of the characteristic
X-rays that are emitted and their intensity to identify and quantify the elements
present in the sample. The following sections describe each of the components
in greater detail.
Radioisotope Sources
An X-ray source will excite characteristic X-rays
from an element only if the source energy is greater than the binding energy,
or absorption edge energy, of the electrons in a given electron shell. A given individual
source can analyze only certain elements.
Analysis is more sensitive for an element with an absorption edge energy similar
to, but less than, the excitation energy of the source. For example, when using
a cadmium-109 (C-109) source, FPXRF would exhibit more sensitivity to zirconium,
which has a K shell energy of 15.7 kiloelectron volts (keV), than for chromium,
which has a K shell energy of 5.41 keV.
The radioisotope sources that are becoming
standard in FPXRF units are Fe-55, Cd-109, and Am-241. Elements that those sources
commonly analyze include:
- Fe-55:
sulfur (S), potassium (K), calcium (Ca), titanium (Ti), and chromium (Cr)
- Cd-109: vanadium (V), Cr, manganese (Mn), iron (Fe), cobalt (Co), nickel (Ni), copper
(Cu), zinc (Zn), arsenic (As), selenium (Se), strontium (Sr), zirconium (Zr),
molybdenum (Mo), mercury (Hg), lead (Pb), rubidium (Rb), and uranium (U)
- Am-241: cadmium (Cd), tin (Sn), antimony (Sb), barium (Ba), and silver (Ag)
Because individual sources by nature reliably
analyze only a limited number of sources, FPXRF instruments that use more than
one source have been developed, allowing them to analyze a greater number and
range of elements. Typical arrangements of such multisource instruments include
Cd-109 and Am- 241 or Fe-55, Cd-109, and Am-241.
X-ray Tube Sources
Miniature x-ray tube sources are now being employed
by a number of vendors. The advantage of the x-ray tube sources is that it does
not require licensing or special shipping, as do XRF units employing radioactive
sources. These units usually have a low-power hot-filament cathode x-ray tube.
The transmission anode operates at a high enough energy range (~35 keV) in order
to simultaneously excite a large range of elements (k through u). Interferences
and sensitivity problems associated with high energy sources are corrected using
sophisticated software built into the XRF unit.
Detectors
Two basic types of detectors are used in FPXRF
units: gas-filled and solid-state.
Each detector has its advantages and limitations and is better suited to some
applications than to others.
Common solid-state detectors include Si(Li), HgI2,
and silicon pin diode. Among those detectors, the Si(Li) is capable of the highest
resolution but is quite temperature-sensitive and will register signal "noise"
if not cooled sufficiently. The Si(Li) has a resolution of 170 electron volts
(eV) if cooled to at least –90°C, either with liquid nitrogen or by thermoelectric
cooling that uses the Peltier
effect. The HgI2 detector can operate at a moderately subambient
temperature and is cooled by use of the Peltier effect. It has a resolution
of 270 to 300 eV. The silicon pin diode detector operates near ambient temperatures
and is cooled only slightly by use of the Peltier effect. It has a resolution
of 250 eV.
Some elements produce peaks that are near
each other in the spectrum, while very high concentrations of one element may
produce a peak that overwhelms the peaks of other elements that are present
at lower concentrations. The higher the resolution, the better the detector
is able to separate characteristic peaks. The XRF operator must be careful to
select an FPXRF unit that has sufficient resolution to satisfy the data quality
needs of the project. The following link provides an illustration of this concept
by providing the resolution differences among some common XRF detectors. Resolution is discussed in greater detail in
a later section.
Mode of Operation
The radioisotope source or sources are housed in
a metal turret, with additional lead shielding inside the probe. To perform
an analysis, a sample is positioned in front of the plastic film measurement
window of the probe and measurement of the sample is initiated, usually by depressing
a trigger or start button. Doing so exposes the sample to the source radiation.
For units that use multiple sources, after the sample has been exposed to one
source, the turret is rotated to expose it to the next source. The length of
time the sample actually is exposed to each source is referred to as the count
time. The sample is exposed to the radioactive source for a number of seconds.
Fluorescent and backscattered X-rays from the sample reenter the analyzer through
the window and are counted by the instrument's detector. X-rays emitted by the
sample at each energy level are called "counts". The detector records
the counts, measures the energy of each X-ray and builds a spectrum of analyte
peaks on a multichannel analyzer (MCA). The unit’s software integrates the peaks
to produce a readout of concentrations of analytes, and, usually, the standard
deviation for each analyte. Numerous sample results and spectra can be stored
for later viewing, downloading into a computer, or printing. Some units also
allow the operator to recall previous results and even to view their spectra.
At the completion of the exposure time, the instrument software statistically
computes a concentration from the readings collected from each energy level
along the spectrum. Count times are not to be confused with the total analytical
time, which includes all of the analytical functions, such as rotation of the
source into position, and processing of the results by the instrument software,
in addition to the count time of each source.
Count times from 30 seconds per source to as long
as 200 seconds per source can be employed, depending on the data quality needs
of the project. As count times increase, the detector collects a larger number
of X-rays from the sample, including more X-rays from elements that are present
at comparatively lower concentrations. For that reason, the longer the count
time, the lower the detection limits; typically, quadrupling the count time
will cut the detection limit in half. For example, if a 50-second count time
yields a detection limit of 100 parts per million (ppm) for a given element,
increasing the count time to 200 seconds will lower the detection limit to approximately
50 PPM Using the instrument's software, the operator can select the appropriate
count times.
An FPXRF detector can be operated in the in situ
or the intrusive mode. Count times of 30 to 60 seconds per source are common
for in situ analysis, while count times for intrusive analysis may be as long
as 200 seconds per source. The particular requirements of the job, such as the
required detection limits or data sample precision, and the purpose of sampling--for
field screening or for definitive analysis--will determine which mode is appropriate
and what count times are needed. Descriptions of each mode follow.
In
situ analysis refers to the rapid screening of soils in place. For in situ operation,
the window of the probe is placed in direct contact with the surface to be analyzed,
and a trigger is pulled, much as one would fire a gun. Because analyses in this
mode typically are completed very quickly (in less than one minute) and sample
heterogeneity of the samples sometimes is a concern, it is recommended that
three to four measurements be taken in a small area and the values be averaged
to determine the concentrations of metals. Intrusive
analysis used to ensure greater precision
when lower detection limits are needed. Those goals are achieved through more
extensive sample preparation and longer analysis times to reduce heterogeneity
among samples and increase the sensitivity of the instrument, respectively.
For intrusive operation, a sample is collected, prepared (usually by homogenizing,
drying, grinding, and sieving), and placed in a 31- or 40-millimeter (mm) polyethylene
sample cup that has a transparent Mylar window. The sample cup is placed over
the probe window (some units provide a safety cover for intrusive analysis)
and analyzed. Some FPXRF instruments can analyze samples in either mode, while
others have only one mode of operation.Standard Operating Procedures (SOPs)
While a clear distinction is made here between
in situ analysis and fully intrusive analysis, sample analysis is in reality
a continuum. Thorough homogenization will improve the precision and accuracy
of the analysis dramatically; an "in situ prepared" sample can be
collected, homogenized, and analyzed right next to the sample location (possibly
right through a plastic bag used for homogenization). Drying the sample also
may improve the results significantly, and, depending on the project’s data
quality objectives, homogenization and drying may be all the preparation required
for an intrusive analysis. Preparation of samples is discussed in greater detail
in a later section.
Target Analytes
The target analytes are metals and
other nonmetallic elements, such as arsenic and selenium.
Performance Specs
Performance specifications include information
about interferences, detection limits, calibration, sample preparation, quality
control, and precision and accuracy.
Interferences
There are a number of factors known as interferences
that can affect the detection and quantification of elements in a sample. Some
interferences can be inherent in the method of analysis, while others are the
result of the instrument's setup, such as calibration methods. Other interferences
may arise from outside sources, such as the sample matrix (for example, soils
and sediment). Some factors can be prevented or minimized through careful preparation
and sample design; others are natural effects that must be taken into consideration.
To produce useful data, it is important that the analyst understand the interferences.
Their effects and the procedures used to evaluate them are described below.
Matrix Effects
Matrix effects can cause a great
deal of variation in sample analyses. Physical matrix effects result from variations
in the physical character of the sample soils, such as particle size, uniformity,
homogeneity, and condition of the surface. The FPXRF demonstration conducted
under the Superfund Innovative Technology Evaluation (SITE) program provided convincing evidence that the
heterogeneity of the sample generally has the greatest effect on comparability
with confirmatory samples. Every effort should be made to homogenize soil samples
thoroughly before analysis. One way to reduce particle size effects is to grind
and sieve all soil samples to a uniform particle size.
Moisture Effects
Moisture content above 20 percent may cause
problems, since moisture alters the soil matrix for which the FPXRF has been
calibrated. This problem can be minimized by drying, preferably in a convection
or toaster oven. Drying by microwave can increase variability between the FPXRF
data and confirmatory data and can cause arcing if fragments of metal are present
in the sample.
Sampling Effects
In environmental samples, typical X-ray penetration
depths range from 0.1 millimeter (mm) to 1 mm. Inconsistent positioning of samples
in front of the probe window is a potential source of error because the X-ray
signal decreases as the distance from the radioactive source increases. Maintaining
a consistent distance between the window and the sample minimizes that problem.
For best results, the window of the probe should be in direct contact with the
sample.
Chemistry Effects
Chemical matrix effects also can occur as
X-ray absorption and enhancement phenomena. For example, iron tends to absorb
copper X-rays, while chromium actually will be enhanced in the presence of iron.
The effects can be corrected mathematically through the FPXRF instrument’s software.
Detector
Resolution Effects
The resolution of the detector
may cause problems in analyzing some elements. If the energy difference between
the characteristic X-rays of two elements (as measured in eV) is less than the
resolution of the detector in eVs, the detector will not be able to resolve
the peaks. In other words, if two peaks are 240 eVs apart, but the resolution
of the detector is 270 eV, the detector will have difficulty in differentiating
those peaks. A common example is the overlap of the arsenic K peak with the
lead L peak. With the use of mathematical corrections that subtract the lead
interference, lead can be measured from the lead L peak and arsenic still can
be measured from the arsenic K peak. However, concentrations of arsenic cannot
be calculated efficiently for samples that have lead to arsenic ratios of 10
to 1 or more, because the lead peak will overwhelm the arsenic peak completely.
Detection Limits
An FPXRF operator must consider two types
of detection limits: instrument detection limits (DL) and method detection limits
(MDL). A DL is the absolute threshold concentration of a given element that
a particular instrument can resolve, as determined by the standard
deviation (SD) of an individual analytical
result. DLs of 10 to 100 PPM are typical for soil samples, although DLs may
be higher for elements like chromium and cadmium that have characteristic X-ray
peaks far removed from the energy level of the sources typically used.
MDLs depend on the analytical method (such as
preparation and analysis times) and may be higher than DLs. The results of replicate
measurements of a low-concentration sample can be used to generate an average
site-specific MDL. The MDL is defined as three times the SD of the results for
a replicate analysis of a low-concentration sample. With the exception of chromium
which has a MDL as high as 900 milligram per kilogram (mg/kg) depending the
instrument being used, the MDLs for most analytes are in the range of 40 to
200 mg/kg.
Click to view a comparison of method
detection limits for six commercial
FPXRF instruments.
Calibration
FPXRF units are calibrated by any of several methods.
The methods will vary according to the make of the unit and the use to which
the data are to be put, such as for screening or for definitive analysis. Basically,
there are two types of calibration, although there is some overlap between two.
Fundamental Parameters Calibration
The fundamental parameters (FP) calibration is a “standardless” calibration. Rather
than calibrating a unit's calibration
curve by measuring its response to standards
that contain analytes of known concentrations, FP calibration relies on the
known physics of the spectrometer's response to pure elements to set the calibration.
Built-in mathematical algorithms are used to adjust the calibration for analysis
of soil samples and to compensate for the effects of the soil matrix. The FP
calibration is performed by the manufacturer, but the analyst can adjust the
calibration curves (slope and y-intercept) on the bases of results of analyses
of check samples, such as standard reference materials (SRM), which are analyzed
in the field.
Empirical Calibration
In performing an empirical
calibration, a number of actual samples,
such as site-specific calibration standards (SSCS), are used, and the instrument's
measurement of the concentrations of known analytes in the samples are measured.
Empirical calibration is effective because the samples used closely match the
sample matrix. SSCSs are well-prepared samples collected from the site of interest
in which the concentrations of analytes have been determined by inductively
coupled plasma (ICP), atomic absorption (AA), or other methods approved by the
US Environmental Protection Agency (EPA). The standards should contain all the
analytes of interest and interfering analytes. Manufacturers recommend that
10 to 20 calibration samples be used to generate a calibration
curve.
Compton Normalization
The Compton
normalization method incorporates elements
of both empirical and FP calibration. A single, well-characterized standard,
such as an SRM or a SSCS, is analyzed, and the data are normalized for the Compton
peak. The Compton peak is produced from incoherent backscattering of X-ray radiation
from the excitation source and is present in the spectrum of every sample. The
intensity of the Compton peak changes as various matrices affect the way in
which source radiation is scattered. For that reason, normalizing to the Compton
peak can reduce problems with matrix effects that vary among samples. Compton
normalization is similar to the use of internal standards in analysis for organic
analytes.
Sample Preparation
Procedures for sample preparation for in situ and
intrusive analysis vary considerably, since the two methods serve completely
different purposes. Sample preparation for in situ analysis is fairly straightforward,
while sample preparation for intrusive analysis can be fairly complicated, depending
on the data quality required.
In situ or
“point-and-shoot” analysis requires little sample preparation.
- First, any unrepresentative debris, such as rocks, pebbles, leaves, vegetation,
roots, and so forth, should be removed from the surface of the soil.
- Second, the surface must be smooth, so that the probe window makes good contact
with the soil surface.
- Last, the surface of the soil should not be saturated to the point that ponded
water is present.
For an "in situ prepared" sample:
- Soil
from the sampling point is collected, and all unrepresentative debris, such
as rocks, pebbles, leaves, vegetation, roots, and so forth, is removed.
- The
soil is thoroughly homogenized.
- The
sample probe is placed directly on the soil for analysis, as with a true
in situ sample, or the sample can be analyzed directly through a plastic
bag used for homogenization.
For intrusive analysis, the sample first must
be collected and then prepared for analysis in a sample cup. Some or all of
the following steps are necessary, depending on the data quality needed:
- The
most important preparation step is thorough homogenization. Mixing the sample
in a plastic bag works well.
- Any
large unrepresentative debris should be removed from the sample.
- If the
sample contains more than 20 percent moisture, the sample should be dried,
preferably in a convection or toaster oven. Drying in a microwave oven is
discouraged because doing so can increase the variability of results and
arcing can occur when metal fragments are present in the sample.
- If a
high degree of precision is required, the sample should be passed through
a sieve. If the sample is not wet (has a moisture content of less than 20
percent) and is not high in clay content, the sample can be sieved in the
field before it is placed in a container. Otherwise, the sample is ground
with a mortar and pestle and passed through a 40- or 60-mesh sieve after
drying.
- Finally,
the sample is placed in a 31- or 40-mm polyethylene cup and covered with
Mylar film.
Quality
Control
Ensuring that the data generated by FPXRF analysis
are of a known quality is vital to ensuring the usefulness of those data, regardless
of their purpose. Quality control (QC) measures take several forms and can be
performed in the field, during sample analysis, and after sample data have been
collected. The amount and type of QC necessary will depend on the project's
data quality objectives. A much higher degree of QC is necessary to produce
defensible, definitive data, but analytical results from intrusive analysis
have been demonstrated to compare favorably with results obtained through traditional
laboratory methods, given that sample preparation has been thorough and QC adequate.
By nature, results obtained in situ are of lower quality because of the lack
of sample preparation, but, with the use of proper QC, in situ data can be corrected.
A typical QC program would include the following measures:
An energy
calibration check sample at least twice daily
An instrument
blank for every 20 environmental samples
A method
blank for every 20 samples
A calibration
verification check sample for every 20 samples
A precision
sample for every 20 environmental samples.
A confirmatory
sample for every 10 environmental samples
Each of the measures identified above is discussed
in detail below.
Energy calibration check samples are used to
test FP calibrations. A check sample consists of a pure element, such as iron,
lead, or copper, and is analyzed to determine whether the characteristic X-ray
lines are shifting, which would indicate drift in the detector. The check also
serves as a gain check in the event that ambient temperatures are fluctuating
significantly (more than 10 to 20° F). The energy calibration check should
be run at a frequency consistent with the manufacturer's recommendations. Generally,
the check would be performed at the beginning of each working day, after the
batteries have been changed or the instrument shut off, at the end of each working
day, and at any other time at which the instrument operator believes that drift
is occurring during analysis.
Two types of blanks can be used during FPXRF analysis.
The first is an instrument blank, which is used to verify that there is no contamination
in the spectrometer or on the probe window. The instrument blank can be silicon
dioxide, a Teflon block, or a quartz block. The instrument blank should be analyzed
a minimum of once daily, preferably once for every 20 samples, and should not
contain any target analytes at levels higher than the MDL. The second type of
blank is a method blank. The method blank is used to monitor sampling and analysis
methods for laboratory-induced contaminants or interferences. The method blank
can be “clean” silica sand or lithium carbonate that undergoes the same sample
preparation procedures as the environmental samples. The method blank should
be analyzed with the same frequency as the instrument blank and should not contain
any target analytes at levels higher than the MDL.
Precision and Accuracy
Calibration verification check samples
are used to check the accuracy of the instrument and assess the stability and
consistency of the analysis of the target analytes. Accuracy is a measure of
the instrument's ability to measure the “true” concentration of an element in
a sample. The check sample can be an SSCS or an SRM, such as the National Institute
of Standards and Technology (NIST)
SRMs, that contains the target analytes,
preferably at concentrations near any action levels for the site. The check
sample should be run at the beginning and the end of each day or for every 20
environmental samples. The percent difference (%D) between the true value and
the measured value should be less than 20 percent.
Instrument precision refers to an instrument's
ability to produce the same result for a number of measurements of the same
sample. The precision of FPXRF measurements is monitored by performing several
analyses of samples that contain low, medium, and high concentrations of target
analytes. It is especially important to know the precision of the instrument
in measuring concentrations that are similar to action levels, because precision
is dependent on analyte concentrations of analytes: as the concentration increases,
the precision increases. A minimum of one precision sample should be run per
day by conducting from 7 to 10 replicate measurements of the sample. The precision
is assessed by calculating a relative standard deviation (RSD) of the replicate
measurements for the analyte. The RSD values should be less than 20 percent
for most analytes, except chromium, for which the value should be less than
30 percent.
Click to see a comparison of instrument precision.
Click
to view the percent
recovery by the FPXRF instrument for a number of metals in performance evaluations
and standard reference materials.
Click
to view the performance of an FPXRF instrument during the analysis of commercial performance
evaluation (PE) samples. PE samples are commercially available standards containing
certified concentrations of various target analytes.
Confirmatory samples are collected from the same
sample material that is analyzed on site, but are sent to an off-site laboratory
for formal analysis. The results of the on-site analysis are compared with the
results of the off-site analysis to determine whether they are comparable within
the acceptable range. The acceptable range is determined by the analytical method,
if applicable, or by the user. The purpose of a confirmatory sample is to judge
the accuracy of the data obtained by analysis on site and to allow corrections,
if necessary. One confirmatory sample usually is submitted for every 10 to 20
samples analyzed on site, depending on the nature of the job.
Advantages
Most instruments weigh less than 30 pounds and
can be operated using battery power for 8 to 10 hours.
A sample can be analyzed in less than five
minutes. Throughput is a measure of the maximum rate of analysis that realistically
can be obtained when using an instrument. That measure includes not only analytical
time, but all sample preparation, QC, and data processing necessary to produce
useable results. Throughput usually is expressed in samples per hour or samples
per day. A throughput of 50 to 100 samples a day typically can be achieved for
intrusive analysis, and as many as 150 samples per day can be analyzed in situ.
Analyses of as many as 35 elements can be performed
simultaneously in a single analysis.
The sample is not destroyed during preparation
or analysis; therefore, it is possible to perform replicate analyses on a sample
and send the same sample for confirmatory analysis, so that comparability studies
can be performed. The sample also can be archived for later use as a soil standard.
Because no solvents or acids are used for sample
extraction, no waste is generated; disposal costs therefore are eliminated.
Operators usually can be trained in one or two
days. The software is menu-driven. No data manipulation is required. Instruments
are marketed for use by general scientists.
Little or no sample preparation is required;
therefore, sample throughput is enhanced and time and money are saved.
Limitations
Detection limits for chromium are 200 mg/kg
or higher. Action levels for some elements, such as arsenic or cadmium, may
be lower than the detection limits of XRF.
Concentrations of elements in different types
of soil or matrices might change, causing interferences--for example, between
arsenic and lead. Site-specific calibration standards can compensate for some
of those effects.
It is difficult to obtain soil standards. One
of the best sources is SRMs from NIST. Those standards cost from $200 to $500
each.
A specific license is required to operate some
FPXRF instruments. The total cost of attending a radiation safety course, obtaining
the necessary paperwork, and paying the fee for the license can range from $500
to $1,000.
The Cd-109 source should be replaced every two
years. The cost of replacement is approximately $4,000 to $5,000.
Any instrument that has a Si(Li) detector will
require liquid nitrogen and a dewar (aluminum container) to hold the liquid
nitrogen. This requirement adds the time and cost of obtaining and handling
liquid nitrogen to cool an instrument with a Si(Li) detector before analysis
can be performed.
Cost
Data
XRF costs vary
significantly. Instrument design and accessories affect instrument prices. Manufacturers
listed below should be contacted directly for cost information.
Additional Resources
Comparing Field Portable X-Ray Fluorescence (XRF) To Laboratory Analysis Of Heavy Metals In Soil
Niton User's
Guide Version 5.0
On-site Analysis
of Metals in Liquids
Sample Handling
Strategies for Accurate Lead-in-Soil Measurements in the Field and Laboratory
Vendor/Instrument Information
| HNU
Systems, Inc. |
SEFA-P
Analyzer |
|
| Scitech Corporation |
AP Spectrum
Analyzer |
|
| TN Spectrace |
TN Pb Analyzer |
|
Verification/Evaluation Reports
Verification of
the performance of site characterization and field analytical technologies is
conducted through a variety of programs. Evaluation and verification reports
from EPA's Superfund Innovative Technologies Evaluation (SITE) Measuring and
Monitoring Program, EPA's Environmental Technology Verification Program (ETV)
program, along with links to certification statements from California EPA's
(CalEPA) California Environmental Technology Certification Program, are provided
below.
Superfund
Innovative Technologies Evaluation (SITE) Measuring and Monitoring Program
The SITE Demonstration Program encourages the development and implementation
of innovative treatment technologies for (1) remediation of hazardous waste
sites and (2) monitoring and measurement. In the SITE Demonstration Program,
the technology is field-tested on hazardous waste materials. Engineering and
cost data on the innovative technologies are gathered so that potential users
can assess the technology's applicability to a particular site. Data collected
during the field demonstration are used to assess the performance of the technology,
the potential need for pre- and post-treatment processing of the waste, applicable
types of wastes and waste matrices, potential operating problems, and approximate
capital and operating costs.
See ETV reports
below
EPA's
Environmental Technology Verification (ETV) Program
EPA's Environmental Technology Verification (ETV) Program verifies the performance
of innovative technologies. ETV was created to substantially accelerate the
entrance of new environmental technologies into the domestic and international
marketplaces. ETV verifies commercialized, private sector technologies. After
the technology has been tested, the companies will receive a verification report
that they can use in marketing their products. The results of the testing also
are available on the Internet. The following reports from the ETV program are
available for x-ray fluorescence:
- HNU
Systems SEFA-P was verified for detection and measurement of a series
of inorganic analytes in soil. The primary target analytes were arsenic,
barium, chromium, copper, lead, and zinc; nickel, iron, cadmium, and
antimony were secondary analytes. The verification documents available
consist of a verification report.
- The
Metorex X-MET 920-P and 940Field Portable X-ray Fluorescence Analyzer
was verified for detection and measurement of a series of inorganic
analytes in soil. The primary target analytes were arsenic, barium,
chromium, copper, lead, and zinc; nickel, iron, cadmium, and antimony
were secondary analytes. The verification documents available consist
of a verification report.
- The
Metorex X-MET 920-MP Fluorescence Analyzer was verified for detection
and measurement of a series of inorganic analytes in soil. The primary
target analytes were arsenic, barium, chromium, copper, lead, and zinc;
nickel, iron, cadmium, and antimony were secondary analytes. The verification
documents available consist of a verification
report.
- The
Niton XL Spectrum Analyzer was verified for detection and measurement
of a series of inorganic analytes in soil. The primary target analytes
were arsenic, barium, chromium, copper, lead, and zinc; nickel, iron,
cadmium, and antimony were secondary analytes. The verification documents
available consist of a verification report.
- The
Scitec MAP Spectrum Analyzer was verified for detection and measurement
of a series of inorganic analytes in soil. The primary target analytes
were arsenic, barium, chromium, copper, lead, and zinc; nickel, iron,
cadmium, and antimony were secondary analytes. The verification documents
available consist of a verification
report.
- The
Spectrace TN 9000 and TN Pb Field Portable X-ray Fluorescence Analyzers
were verified for detection and measurement of a series of inorganic
analytes in soil. The primary target analytes were arsenic, barium,
chromium, copper, lead, and zinc; nickel, iron, cadmium, and antimony
were secondary analytes. The verification documents available consist
of a verification report.
California
EPA's California Environmental Technology Certification Program
CalEPA's environmental technology certification program
is a voluntary program that provides participating technology developers, manufacturers,
and vendors an independent, recognized third-party evaluation of the performance
of new and mature environmental technologies. Developers and manufacturers define
quantitative performance claims for their technologies and provide supporting
documentation; CalEPA reviews that information and, when necessary, conducts
additional testing to verify the claims. The technologies, equipment, and products
that are proven to work as claimed are given official state certification. The
certification program is voluntary and self-supporting. Companies participating
in the program pay the costs of the evaluation and certification of their technologies.
Technologies that
have been certified through this program are listed below. Links are provided
to the web sites that provide the Certified Environmental Technology Transfer
Advisory and Certification Notice for the technologies.
No
reports available for this technology
Disclaimer
http://www.clu-in.org/char/technologies/xrf.cfm
Page Last Modified: May 18, 2007

|